Puritan Bennett™ 840 Ventilator Settings

  • Ideal body weight (IBW): 0.3 to 7.0 kg with NeoMode 2.0 (0.66 to 15 lbs), 7.0 to 24 kg (15 to 53 lbs), 25 to 150 kg (55 to 330 lbs)
  • Modes: Assist/Control (A/C), synchronized intermittent mandatory ventilation (SIMV), or spontaneous (SPONT), Bi-Level
  • Software Options: NeoMode 2.0, Leak Compensation, Tube Compensation, Bi-Level, Volume Ventilation Plus (Volume Control Plus and Volume Support), Proportional Assist™* Ventilation Plus (PAV™*+), Respiratory Mechanics and Trending
  • Mandatory breath types: Volume Control (VC), Pressure Control (PC) or Volume Control Plus with Volume Ventilation Plus option
  • Spontaneous breath types: Pressure Supported (PS), Volume Supported (VS), Proportional Assist (PA), none
  • Vent type: Invasive or Noninvasive
  • Pressure support (PSUPP): 0 to 70 cm H2O
  • Rise time %: 1% to 100%
  • Expiratory sensitivity (ESENS): 1% to 80%; 1 L/min to 10 L/min with PAV™*+
  • Tidal volume (VT): 25 to 2,500 mL, 2 to 315 mL with NeoMode
  • Respiratory rate (f): 1.0 to 100 /min, 1 to 150 /min with NeoMode
  • Peak inspiratory flow (VMAX): 3 to 150 L/min for IBW > 24 kg; 3 to 60 L/min for IBW ≤ 24 kg, 1 to 30 L/min with NeoMode
  • Flow pattern: Square or descending ramp
  • Plateau time (TPL): 0.0 to 2.0 seconds
  • Inspiratory pressure (PI): 5 to 90 cm H2O
  • Constant during rate change: Inspiratory time (TI), I:E ratio or expiratory
  • time (TE)
  • Inspiratory time (TI): 0.2 to 8.0 seconds
  • I:E ratio: ≤ 1:299-4.00:1
  • Expiratory time (TE): TE ≥ 0.2 second
  • Trigger type: Pressure (PTRIG) or flow (VTRIG Flow-by flow triggering)
  • Pressure sensitivity (PSENS): 0.1 to 20 cm H2O below PEEP
  • Flow sensitivity (VSENS): 0.2 to 20 L/min, 0.1 to 10 L/min with NeoMode
  • O2%: 21% to 100%
  • PEEP: 0 to 45 cm H2O
  • Apnea ventilation: Apnea mandatory type-volume control (VC) or pressure control (PC)
  • Apnea flow pattern: Square or descending ramp
  • Apnea peak flow (VMAX): ≥ 1.0 L/min to ≤ 30 L/min, ≥ 3.0 L/min to ≤ 60 L/min, ≥ 3.0 L/min to ≤ 150 L/min
  • Apnea tidal volume (VT): 3 mL to 315 mL for neonatal; 25 mL to 2500 mL for ped/adult
  • Apnea inspiratory pressure (PI): 5 to 90 cm H2O
  • Apnea inspiratory time (TI): 0.2 to 8.0 seconds
  • Apnea interval (TA): 10 to 60 seconds
  • Apnea respiratory rate (f): 2.0 to 40 1/min
  • Apnea O2%: 21% to 100%
  • Apnea expiratory time (TE): ≥ 0.2 seconds
  • Disconnect sensitivity (DSENS): 20% to 95% or OFF in NIV
  • Humidification type: Heat-moisture exchanger (HME), nonheated expiratory tube or heated expiratory tube
  • Humidification volume: 100 to 1,000 mL
  • Patient circuit type: Pediatric, adult or neonate with NeoMode option