Technical data Evita 2 dura

Patient type Adults and children (body weight of at least 3 kg (6.6 lbs))
With NeoFlow option: infants and neonates (min. body weight of 0.5 kg (1.1 lbs))
Ventilation settings
Ventilation mode • IPPV, IPPVAssist (CMV, CMVAssist)
• SIMV, SIMVASB (SIMV, SIMV/Psupp)
• MMV, MMVASB (MMV, MMV/Psupp)
• BIPAP1), BIPAP1)ASB, BIPAP1)Assist (PCV+, PCV+/Psupp, PCV+Assist)
• CPAP, CPAPASB (CPAP, CPAP/Psupp)
• APRV (optional)
• ILV (optional)
Enhancements • AutoFlowTM – Automatic adaptation of inspiratory flow in volume
controlled modes
• ATCTM – Automatic Tube Compensation (optional)
• NIV – Mask Ventilation (optional)
Ventilation frequency (f) 0 to 100 /min, 0 to150 /min (Neonatal)
Inspiration time (Tinsp) 0.1 to 10 s
Tidal volume (VT) (BTPS*) • 0.1 to 2.0 L (Adult)
• 0.02 to 0.3 L (Pediatric)
• 0.003 to 0.1 L (Neonatal)
Inspiratory flow • 6 to 120 L/min (Adult)
• 6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure 0 to 80 mbar (cmH2O)
PEEP / intermittent PEEP 0 to 35 mbar (cmH2O)
Pressureassist (PASB) (Psupp) 0 to 80 mbar (cmH2O)
Rise time for inspiratory pressure 0 to 2 s
O2 concentration 21 to 100 Vol.%
Multi-sense Trigger Criteria Internal automatic pressure trigger, Flow, Volume
(Flow adjustable 0.3 to 15 L/min)
Measured values displayed
Airway pressure Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure
(0 to 99 mbar/cmH2O)
Minute volume (MV), (BTPS*) MV, MVspont, MVleak (0 to 99 L/min)
Tidal volume (VT), (BTPS*) Inspired VT, expired VT, VTASB (VTPS) (0 to 3999 mL)
Breathing frequency (f) ftotal, fspon, fmand. (0 to 300/bpm)
O2 concentration (FiO2) Inspired O2 concentration (15 to 100 Vol.%)
Lung mechanics • Resistance ((0.0 to 600 mbar/L/s)(cmH2O/L/s))
• Compliance ((0.0 to 300 mL/mbar)(mL/cmH2O))
Breathing gas temperature 18° to 51°C
Waveforms Airway pressure-time, flow-time, volume-time, …
Trends (optional) FiO2, MV, VT, f, PEEPi, R, C, etCO2, …
Loops (optional) Paw-V, V-Flow, Flow-Paw, …
Capnography (etCO2) (optional) • 0 to 100 mmHg
CO2 production (VCO2) • 0 to 999 mL/min, STPD*
Serial dead space Vds • 0 to 999 mL, BTPS*
Dead space ventilation (Vds/VT) • 0 to 99%
Alarms / Monitoring
Airway pressure High/Low
Expired minute volume High/Low
Tidal volume High
Apnea alarm time 5 to 60 s
Spontaneous breath frequency High
Inspired O2 concentration High/Low
Breathing gas temperature High
SpO2 pulse (optional) High/Low
etCO2 (optional) High/Low
Performance data
Max. flow for pressure support
and spontaneous breathing 180 L/min (adult), 60 L/min (pediatric)
Valve response time T0…90 ≤ 5 ms
Control principle Time cycled, volume constant, pressure-controlled
Safety relief valve 100 mbar (cmH2O)
Leakage compensation
Hose system compensation
Outlet for pneumatic nebulizer
Operating data
Mains power connection 100 to 240 V, 50/60 Hz, 10 to 30 V DC (optional)
Power consumption Approx. 125 W
Gas supply operating pressure O2, air: 2.7 to 6 bar/39 to 87 PSI
Physical Specifications
Dimensions ventilator (WxHxD) 530x290x450 cm (20.9×11.4×17.7 inches) (without trolley)
Weight basic unit Approx. 27 kg (60 lbs.)
Diagonal screen size 6.5″ TFT LCD color display
Machine outputs
Digital output Output and reception via an RS 232 C interface
Digital output (optional) For output and reception via two RS 232 C interfaces
Analog output (optional) For analog output of two measured values